Performs daily empty‑chamber test to assess air‑removal and steam‑penetration capability of pre‑vacuum steam sterilizers.
|
Ref.# |
1A8199 |
|
Specification |
115*125mm,100PCS/CTN |
|
Description |
This product is not required when using standard testpacks. Place the test pack near the sterilizer exhaust vent and run a 134"C cycle for 3.5minutes. |
|
Scope of use |
Applicable for testing the air removal and steam penetration performance of pre-vacuum pressure steam sterilizers at 134°C for 3.5 minutes. |
|
Perservation Method |
During storage and use, test packs must avoid contact with acids/alkalis, maintain dry conditions (RH <5096), and be shielded from UV, fluorescent, and sunlight. |